BD Alaris EtCO2 Module 8300
The BD Alaris EtCO2 Module 8300 provides continuous respiratory monitoring by measuring CO2 levels, including exhaled EtCO2, inhaled FiCO2, and respiratory rate. This side-stream capnography system draws a small gas sample—about 15 microliters—through a micro-sample cell from either a ventilator disposable or directly from the patient. Moisture and patient secretions are effectively filtered out by the microstream inline filter, ensuring that the CO2 waveform remains intact. This module is designed to enhance patient safety by reducing risks associated with opioid infusions and to offer flexible, customizable profiles that work across different patient care areas. Its sampling lines eliminate the need for moving parts, external sensors, or frequent calibration, and it supports a wide range of sampling needs, whether oxygen is being delivered or not.
Key Features:
- Continuous Respiratory Monitoring:
- Measures exhaled EtCO2, inhaled FiCO2, and respiratory rate.
- Side-Stream Capnography:
- Delivers a precise 15-microliter gas sample through a micro-sample cell.
- Effective Moisture & Secretion Management:
- Uses a microstream inline filter to extract moisture and secretions, preserving the CO2 waveform.
- Enhanced Safety:
- Reduces risks during opioid infusions with customizable profiles tailored to various care settings.
- Simplified Design:
- Utilizes sampling lines that eliminate the need for moving parts, external sensors, and routine calibration.
- Flexible Sampling:
- Supports a broad spectrum of EtCO2 sampling products, with or without oxygen delivery.
Specifications:
-
Dimensions:
- Height: 8.8 in
- Width: 6.9 in
- Depth: 9 in
- Weight: 7.2 lbs
-
Environmental Operating Conditions:
- Atmospheric Pressure: 525 – 4560 mmHg (700 – 6080 hPa)
- Relative Humidity: 20 – 90% (noncondensing)
- Temperature Range: 41 – 104°F (5 – 40°C)
-
Environmental Storage / Transportation Conditions:
- Atmospheric Pressure: 375 – 760 mmHg (500 – 1013 hPa)
- Relative Humidity: 5 – 85% (noncondensing)
- Temperature Range: -4 – 140°F (-20 – 160°C)
-
Flow Rate:
- Nominally 50 mL/min (with a tolerance of -7.5 mL/min to +15 mL/min)
-
Warm-Up Time:
- Approximately 30 seconds
-
Power Requirements:
- Input: 100 – 240V AC, 50/60 Hz, 150 VA max
-
Internal Power Source:
- Battery-operated with an operating time of approximately 5.5 hours
-
Measurement Range:
- EtCO2: 0-99 mmHg
- FiCO2: 0-99 mmHg
- Respiratory Rate: 0-150 bpm