ConMed Hyfrecator 2000 Electrosurgical Unit
The ConMed Hyfrecator 2000 is a versatile electrosurgical unit designed for a wide range of procedures including dermatology, gynecology, ophthalmology, and urology. Built with dual microprocessor technology, this unit delivers precise and accurate power—whether you need full power for broad surface coagulation or delicate, low-power settings in 1/10-watt increments for facial procedures. It is engineered for both high performance and patient safety, and is fully compatible with existing Hyfrecator accessories such as bipolar forceps and reusable electrodes.
Key Features
- Dual Microprocessor Technology:
- Provides unsurpassed output power accuracy and patient safety by reacting quickly to tissue impedance variations.
- Versatile Energy Delivery:
- Offers a wide power range, from low power in 1/10-watt increments for delicate procedures to full power for effective coagulation on larger tissue areas.
- State-of-the-Art Electronic Circuitry:
- Ensures precise and consistent energy delivery across a variety of surgical applications.
- Built-In Self-Diagnostic Testing:
- Automatically runs internal tests at power-on to confirm proper functionality.
- Ergonomic & Lightweight Design:
- Weighs only 6 lbs, making it easy to move and integrate into any operating room setup.
- Compatibility:
- Works seamlessly with your current Hyfrecator accessories, ensuring an easy upgrade from previous models.
Detailed Specifications
Dimensions & Weight
- Height: 7.5 in (19 cm)
- Width: 8.75 in (22.2 cm)
- Depth: 4 in (10.2 cm)
- Weight: 6 lbs (2.7 kg)
Power Accuracy & Output
- Low and Bipolar Modes (LO & BI):
- 0 to 10 W: Accuracy is the greater of 90 mW or 10% of the setting.
- 10 W to maximum: Accuracy is 10% of the setting.
- High Mode (HI):
- Accuracy is the greater of 1 W or 10% of the setting.
- RF Frequency:
- 450 kHz ±50 kHz (open circuit)
- Output Wave Shape:
- Clipped, damped sinusoidal
- Line Regulation:
- ±5% at maximum power for a ±10% line voltage change
- Duty Cycle:
- Intermittent: 30 sec. on, 30 sec. off
- Audio Output:
- Adjustable from >45 dBA to 65 dBA at 1 meter
- 60 Hz Leakage:
- Maximum of 50 µA when the third wire of the line cord is open
Display & Controls
- Output Power Settings:
- Lower Output Terminal: 0-10 W in 0.1 W increments
- Bipolar & High Output Terminals: 0-35 W in 1 W increments
- Functional Modes:
- Hi Mode: Maximum power setting up to 35 W with a peak voltage of 8000 V at a rated load of 1000 ohms, repetition frequency of 24.4 kHz.
- LO Mode: Up to 20 W, maximum voltage 3000 V at a rated load of 500 ohms, repetition frequency of 32.3 kHz.
- BI Mode: Also up to 35 W, with the same voltage and repetition frequency as LO mode.
Environmental Conditions
- Operating Temperature: 10°C to 40°C
- Relative Humidity: 10% to 95% (non-condensing)
- Storage Conditions: -40°C to 70°C, 95% humidity non-condensing
- Operating Altitude: Between 60 meters below and 3000 meters above mean sea level