The B Braun Perfusor Space Infusion Pump is designed for advanced delivery in places like operating rooms, adult care, pediatrics, and emergency settings where precise medication administration is really critical. It’s super compact and lightweight—about 9” x 6” and just 3 pounds—so it’s easy to move around even in crowded care areas.
This pump is easy to use with an intuitive interface backed by a 1,500-entry drug library that’s color-coded for quick selection. The keys give off audible clicks and each alarm is both loud and clearly visible with different tones, making sure you never miss an alert. Plus, it comes with a sturdy yet lightweight pole clamp that can hold up to three pumps mounted either horizontally or vertically.
Specifications:
-
Physical Dimensions:
- Width: 9.8 in (24.9 cm)
- Height: 2.6 in (6.8 cm)
- Depth: 5.9 in (15.2 cm)
- Weight: 1.08 lbs (1.4 kg)
-
Battery:
- Type: NiMH (Rechargeable)
- Operating Time: Approx. 8 Hours at 25 mL/h
- Recharge Time: Approx. 6 Hours
-
Environmental Conditions:
- Operating Conditions:
- Relative Humidity: 30% – 90% (Without condensation)
- Temperature: 41°F – 104°F (5°C – 40°C)
- Atmospheric Pressure: 500 – 1060 bar
- Storage Conditions:
- Relative Humidity: 30% – 90% (Without condensation)
- Temperature: -20°F – 131°F (-20°C – 55°C)
- Atmospheric Pressure: 500 – 1060 bar
- Operating Conditions:
-
Volume Preselection:
- 0.1 – 99.99 mL in increments of 0.01 mL
- 100.0 – 999.0 mL in increments of 0.1 mL
- 1000 – 9999 mL in increments of 1 mL
-
Time Preselection:
- 00:01 – 99:59 hours
-
Accuracy of Set Delivery Rate:
- ± 2% according to IEC/EN 60601-2-24
-
Occlusion Alarm Pressure:
- 9 levels up to 1.2 bar
-
Alarm for Incorrect Dose:
- If an incorrect dose of 0.1 mL occurs due to a device malfunction, the pump will automatically shut off.
-
External Power Supply:
- Rated Voltage: 100 – 240 V AC, 50/60 Hz
- External Low Voltage: 11 – 16 V DC via Connection Lead SP 12 V or via SpaceStation