Baxter 6301 Infusion Pump
The Baxter 6301 is a dual channel volumetric infusion pump designed for precise fluid delivery to two patients at once. It features two independent linear peristaltic pump heads, each capable of being programmed separately so that different solutions can be infused simultaneously. With incremental flow rate settings ranging from 1 to 99.9 mL/hr (in 0.1 mL/hr increments), this pump provides the flexibility needed for a wide variety of clinical applications. Additional features like flow check occlusion alarms, flow rate calculations, and programmed delivery profiles ensure that therapy is both safe and consistent. The system also includes an automatic restart function once an occlusion clears and supports automatic piggybacking for secondary medications.
The pump’s battery performance is robust: when running one channel at rates from 1 to 1400 mL/hr, it lasts approximately 6 hours; if both channels are active at the same rates, the battery life is about 4 hours. This reliability, coupled with its advanced safety features, makes the Baxter 6301 a trusted choice for infusion therapy.
Specifications:
-
Dimensions:
- Height: 11.4 in (29 cm)
- Depth: 5.1 in (13 cm)
- Width: 13 in (32.5 cm)
- Weight: 8.1 kg (11.6 lbs)
-
Battery:
- Type: 12 V, 3.2 Ah sealed lead acid
- Battery Life:
- Approximately 6 hours with one pump running (1 to 1400 mL/hr)
- Approximately 4 hours with both pumps running (1 to 1400 mL/hr)
- 8 hours for more than 80% recharge
-
AC Power Requirements:
- Input: 110/120 V, 60 Hz
- Power Consumption: 50 W
- Power Cord: 2.9 m (9.5 ft) long, Hospital Grade plug
- Fuse: 0.8 A, 250 V, SB, 5.2 mm x 20 mm
-
Flow Rate Range:
- Primary Program: 1 – 999 mL/hr in 1 mL/hr increments and 1 – 99.9 mL/hr in 0.1 mL/hr increments
- Secondary Program: 1 – 999 mL/hr in 1 mL/hr increments and 1 – 99.9 mL/hr in 0.1 mL/hr increments
- (Upper limits can be reduced by authorized service personnel)
-
VTBI Range:
- 1.0 – 99.9 mL in 0.1 mL increments for both primary and secondary channels, with a total volume range of 1 – 9,999 mL
- (Upper limits adjustable by authorized service personnel)
-
Leakage Current: Less than 50 µA (per UL 544)
-
Air-in-line Detection:
- Factory setting (NORM) alarms for air bubbles of approximately 75 µL or larger; setting can be adjusted to MIN, which alarms at around 50 µL or larger
-
Additional Features:
- Automatic restart after occlusion clears
- Automatic piggybacking for secondary medications
Reviews