Curlin 4000 Plus Infusion Pump
The Curlin 4000 Plus Clinical Management System is a multi-therapy ambulatory electronic infusion pump renowned for its low risk of medication errors. This advanced pump supports a variety of infusion modes—including continuous, PCA (patient-controlled analgesia), PCEA (patient-controlled epidural analgesia), subcutaneous, YPN, intermittent, and variable modes—making it ideal for a broad range of clinical applications. It continuously monitors and displays line pressure to help detect potential catheter issues such as occlusion, obstruction, or pinched-off lines, ensuring safe and effective therapy. One of its standout features is an exceptional low flow resolution of just 2 micrometers, which is significantly better than that of other ambulatory pumps.
Key Modes and Features:
-
Continuous Infusion:
Delivers a constant, programmed rate of infusion for steady fluid delivery. -
Patient-Controlled Analgesia (PCA):
Supports therapies requiring a continuous infusion rate along with patient-controlled bolus doses, ensuring timely pain relief. -
TPN with Automatic Ramping:
Provides a consistent infusion rate for parenteral nutritional products with options for tapering at the beginning, end, or both, and includes an early ramp-down feature. -
Intermittent Delivery:
Delivers infusate in programmed intervals and rates, with the option to infuse small amounts between doses to keep the patient’s access site patent. -
Variable Program:
Allows the delivery of varying amounts, rates, and times across up to twenty-four distinct programs to meet complex clinical needs.
Specifications:
-
Dimensions & Weight:
- Length: 5 in
- Width: 4 in
- Depth: 2.5 in
- Weight: 17.5 oz
-
Power Requirements:
- Power Sources: Operates on 2 “C” size alkaline batteries, with the option for AC power or an external rechargeable battery pack.
- Battery Life:
- Approximately 10 hours at 400 mL/hr
- Approximately 30 hours at 125 mL/hr
- Approximately 85 hours at 2 mL/hr
-
General Performance:
- Accuracy: ±6%
- Flow Rate Range: 0.1 mL to 400 mL/hr
- Volume Limits: 1.0 mL to 9,999 mL
- KVO (Keep Vein Open) Rates: 0.0 to 10 mL/hr (default is 0.1 mL/hr)
Additional Benefits:
-
Flexible:
Offers versatile application capabilities to meet specific workflow needs, with access to actionable reports and detailed infusion data that enhance overall performance. -
Safe:
Delivers medications at low flow rates with a resolution up to 25 times better than competing ambulatory pumps, working synergistically with a PC to reduce programming errors and improve outcomes. -
Easy-to-Use:
Features menu-driven protocols and context-sensitive HELP screens, making the device extremely user-friendly and reducing training time. -
Cost-Effective:
The pump’s simplicity, reliable technology, and use of low-cost PVC administration sets help reduce overall ownership costs through lower training and purchasing expenses.