ERBE Erbotom ICC 350 Electrosurgical Unit
The ERBE Erbotom ICC 350 Electrosurgical Unit represents the pinnacle of high-frequency surgical technology, delivering unparalleled performance across a wide range of surgical specialties including general surgery, dermatology, endoscopy, gastroenterology, vascular surgery, gynecology, neurosurgery, plastic surgery, and oral surgery. Developed in close collaboration with leading surgeons, the ICC 350 offers one-of-a-kind functionality with advanced cutting and coagulation modes, ensuring precise tissue effects and improved patient outcomes.
Key Features
- Versatile Surgical Applications:
- Designed for use in diverse fields such as general surgery, GI, vascular, neurosurgery, plastic, and oral surgery.
- Advanced Cutting Modes:
- Monopolar AUTO CUT: Automatically controls voltage for consistent cutting.
- Monopolar HIGH CUT: Features automatic arc control for underwater or fatty tissue cuts.
- Monopolar ENDO CUT: Optimized for endoscopic procedures.
- Multiple Coagulation Modes:
- Monopolar SOFT COAG: Provides precisely controlled coagulation with minimal carbonization at 300 kHz.
- Monopolar FORCED COAG: Uses pulse-modulated current at 1 MHz for rapid hemostasis.
- Monopolar SPRAY COAG: Offers efficient surface coagulation via pulse modulation.
- Bipolar COAG: Ensures safe coagulation with lower voltage levels.
- Intelligent Power Control:
- HF power adjustment in 1-watt increments, with a 7-segment display (3 decimal places) for precise control.
- Integrated Power Peak System (PPS) supports initial incisions, boosting output up to 400 watts at 500 ohms in AUTO CUT mode.
- Robust Safety Measures:
- Incorporates the NESSY Neutral Electrode Safety System and comprehensive self-checks to monitor LF and HF leakage currents.
Detailed Specifications
Dimensions & Weight
- Dimensions (W x H x D):
- 410 mm × 152 mm × 368 mm
- Weight:
- 10.0 kg
Electrical Specifications
- Rated Power Voltage:
- 240V / 230V / 115V / 110V / 100V ±10%
- Frequency:
- 50/60 Hz
- Power Current:
- 4.0 A at 230–240 V; 8.0 A at 100–115 V
- Standby Power Consumption:
- 25 Watts; current: 150 mA at 230–240 V / 300 mA at 100–115 V
- Maximum HF Power Consumption:
- 620 Watts (920 VA)
- Fuses:
- 2 slow-burn fuses: 4 A (230–240 V) / 8 A (100–115 V)
- Potential Equalization Terminal:
- Yes
Cutting & Coagulation Parameters
- Monopolar Cutting:
- AUTO CUT & HIGH CUT:
- Nominal HF output: 300 watts at 500 ohms; Peak HF power with PPS: 400 watts at 500 ohms; Rated frequency: 330 kHz; Sine-shaped voltage waveform.
- ENDO CUT:
- Operates at 200 watts for endoscopic applications.
- AUTO CUT & HIGH CUT:
- Coagulation Modes:
- Soft Coagulation:
- Nominal HF output: 120 watts at 125 ohms; Rated frequency: 330 kHz; Peak voltage up to 1900 Vp.
- Forced Coagulation:
- Nominal HF output: 120 watts at 350 ohms; Rated frequency: 1 MHz; Peak voltage up to 2300 Vp.
- Bipolar Coagulation:
- Nominal HF output: 120 watts at 125 ohms; Voltage waveform is sine-shaped.
- Soft Coagulation:
- HF Power Adjustment:
- Configurable in 1-watt increments via up/down keys, displayed on a 7-segment display with a precision of ±1 digit or ±15%.
Safety & Environmental Conditions
- Safety Class:
- Class I, Type CF (DIN IEC 601-1)
- LF Leakage Current (Self-Check):
- Icritical: 0.05 mA (with capacitive grounding)
- HF Leakage Current (Self-Check):
- Icritical: 150 mA/300 mA
- Storage Conditions:
- Temperature: -40°C to +70°C; Relative Humidity: 30% to 95%
- Operating Conditions:
- Temperature: +10°C to +40°C; Relative Humidity: 30% to 75% (non-condensing)
Reviews