Hospira Plum 360 Infusion Pump
The Hospira Plum 360 is a high-performance, large-volume infusion pump known for its exceptional safety and reliability. Designed to meet the demanding needs of hospitals, this pump features a dual-line cassette system that not only delivers fluid with precision but also automatically removes air from the line through a secondary port—so you won’t need to disturb the patient’s infusion for air clearance. With a built-in secondary delivery system, the Plum 360 minimizes programming errors and ensures that bolus doses are delivered rapidly when needed.
This versatile system is fully compatible with ICU Medical MedNet software, enabling customizable drug libraries that can be tailored to your facility’s specific protocols. The pump communicates effortlessly with other systems on your network through either Ethernet or wireless connections, supporting full IV-EHR interoperability. Additionally, its advanced PlumSet cassette technology integrates free-flow protection and automated concurrent as well as piggyback delivery, while the secondary line connects directly to the cassette for seamless operation.
Specifications:
-
Dimensions:
- Height: 8 in (20 cm)
- Width: 8 in (20 cm)
- Depth: 6 in (15 cm)
- Weight: 10 lb (4.5 kg)
-
Battery:
- Type: Sealed lead-acid rechargeable 6V internal battery
- Operating Time: Approximately 7 hours at 25 mL/hr; about 4 hours at 999 mL/hr
- Recharge Time: Up to 8 hours when operating at 125 mL/hr on one line
-
VTBI Range:
- 0.1 to 99.9 mL (in 0.1 mL increments)
- 100 to 9,999 mL (in 1 mL increments)
-
Delivery Rate Range:
- For Lines A and B: 0.1 to 99.9 mL/hr (in 0.1 mL/hr increments) and 100 to 999 mL/hr (in 1 mL/hr increments)
- Concurrent Delivery: Minimum of 0.5 mL/hr per line
- PlumSet Cumulative: Up to 500 mL/hr (A+B) maximum
- KVO: 1.0 mL/hr or the last delivery rate on the associated line, whichever is less
-
Bolus Delivery (Software Version 15.1 and Later):
- Rates: 1 to 99.9 mL/hr (in 0.1 mL/hr increments) and 100 to 999 mL/hr (in 1 mL/hr increments)
- Maximum Programmable Duration: 1500:00 (hh:mm)
-
Air-in-Line Alarm:
- PlumSet (Distal): Air bolus detection at 0.1 mL or larger; cumulative threshold of 0.25 mL out of 4.9 mL
- PlumSet (Proximal): Air bolus detection at 0.5 mL; cumulative threshold of 1.0 mL
-
Delivery Accuracy:
- 0.1 to 0.9 mL/hr: ±10% (in 0.1 mL/hr increments)
- 1 to 999 mL/hr: ±5% (in 1 mL/hr increments)
-
Additional Parameters:
- Ambient and Fluid Temperature: 22°C ±5°C
- Back Pressure: 0 psi (0 mmHg)
- IV Fluid: Sterile water
- Filling Head Height: 12 to 24 in (30.5 to 61 cm)
Reviews
There are no reviews yet.