Hospira Plum A+3 Infusion Pump
The Hospira Plum A+3 is a triple-channel infusion pump that allows multiple infusion lines to be run simultaneously. Designed for high-demand environments such as ERs, ICUs, and Oncology centers, this advanced pump integrates Hospira MedNet safety software and a barcode point-of-care system to ensure accurate, secure medication delivery. With its wireless-enabled system, clinicians can program and monitor infusions in real time, whether delivering primary, secondary, or concurrent medications in a tiered sequence.
Key Features:
-
Multi-Channel Flexibility:
- Triple-channel design supports simultaneous infusions on up to 6 concurrent medication lines.
- Allows for automated piggyback and concurrent delivery, as well as programmable standby settings, multi-step delivery, loading dose automation, and delayed starts.
-
PlumSet™ Technology:
- Unique technology that delivers infusions with confidence by ensuring consistent dosing.
-
Automated Second-Line Delivery:
- Eliminates the need to manually raise and lower infusion containers, enhancing ease-of-use and efficiency.
-
Air-in-Line Detection & Elimination:
- Built-in air trap captures up to 2 mL of air.
- Automated back-priming removes accumulated air without disconnecting the patient line, reducing the risk of contamination.
-
Versatile Infusion Options:
- Delivers secondary infusions via IV container or syringe.
- Supports a wide variety of programming options to match clinical needs.
-
Wireless Connectivity:
- Facilitates real-time infusion monitoring and supports real-time 5-rights verification.
-
User-Centric Design:
- Intuitive programming screens, clear numeric keypad, and confirmation prompts ensure accurate dosing.
- Integrated system works seamlessly across the entire spectrum of clinical care.
Specifications:
-
Dimensions:
- Height: 19 in
- Width: 15 in
- Depth: 14 in
- Weight: 28 lb (with 3 batteries)
-
Electrical:
- Power Requirement: 120 V AC, 50–60 Hz, 120 VA
- Fuses: F1, F2, 250 V, 0.5 A (internal)
- Power Cord: Hospital-grade AC cord, 10 ft long, with transparent plug and retainer plate
- Battery: Three sealed, lead-acid, rechargeable 6 V batteries (internal)
- Battery Life:
- Minimum 6 hours at ≤125 mL/hr on one line, or deliver 500 mL at ≥126 mL/hr on one line
- Recharge: Approximately 6 hours with the device operating at 125 mL/hr on one line
- Leakage: Meets ANSI/AAMI ES1-1993 safe current limits
- Circuitry Ratings: Voltage: 30 VDC max; Current: 0.25 A max; Contact Rating: 3 Watts max
-
Environment:
- Operating Temperature: 5° to 40° C
- Storage Temperature: -20° to 60° C
- Atmospheric Pressure: 0 – 10,000 ft (0 – 3,000 m) or equivalent
- Relative Humidity: 10 – 90% (max 40° C)
-
General Delivery Parameters:
- Volume Range: 0.1 to 99.9 mL (in 0.1 mL increments); 100 to 9,999 mL (in 1 mL increments)
- Delivery Rate Accuracy: ±5% (1–999 mL/hr), ±10% (0.1–0.9 mL/hr)
- Primary Delivery Rate: 0.1 mL/hr minimum, up to 999 mL/hr maximum
- Secondary Delivery Rate: 0.1 mL/hr minimum, up to 999 mL/hr maximum
- Concurrent Delivery: Minimum 0.5 mL/hr per line, cumulative maximum of 500 mL/hr
- KVO Rate: 1.0 mL/hr or the last primary delivery rate, whichever is less
-
Delivery Units:
- Options include mL/hr, mcg/kg/min, mcg/min, mcg/kg/hr, cg/hr, mg/min, mg/kg/hr, mg/hr, ng/kg/min, g/hr, mEq/hr, Million units/hr, units/min, units/hr, units/kg/hr, units/kg/min, mmol/min, mmol/hr
-
Alarms:
- System alarms include Air-in-Line, Air-in-Line Backpriming, Low Battery, Occlusion, Turn to Run, Flow Detector, VTBI Complete or Dose End
-
Occlusion Alarm Pressure:
- 1–15 psig (default 6 psig)