Medfusion 4000 Wireless Syringe Pump
The Medfusion 4000 Wireless Syringe Pump is the next-generation evolution of the popular Medfusion 3500, renowned for its precise medication delivery in critical care settings—including neonatal and pediatric ICUs. This advanced, customizable pump is designed to help prevent medication errors by streamlining the exchange of delivery information. It automatically detects various syringe types and sizes (from 1 mL to 60 mL) to ensure the correct volume is delivered every time. Integrated with Pharmguard medication security and comprehensive medication libraries, the Medfusion 4000 makes it easy for healthcare providers to administer the right dose safely.
Key Features:
- High Precision Delivery:
- Delivers infusion volumes with an accuracy of ±2%, at flow rates ranging from 0.01 mL/hr to 1130 mL/hr.
- Automatic Syringe Detection:
- Automatically senses a wide variety of syringe types and sizes (1 mL to 60 mL) for optimized performance.
- Wireless Connectivity:
- Connects seamlessly to Pharmguard software, allowing real-time data exchange without interrupting pump operation.
- Smart Pump Technology:
- Reduces medication delivery errors and enhances patient safety with integrated flow sensors that monitor line pressure.
- Customizable System:
- Offers customizable settings and medication libraries to meet specific clinical protocols and prevent errors.
Specifications:
-
Dimensions:
- Height: 6.2 in (16 cm)
- Width: 10.5 in (27 cm)
- Depth: 5.7 in (14.5 cm)
- Weight: 4.54 lbs (2.04 kg)
-
Electrical:
- Internal Battery: Rechargeable lithium-ion
- AC Power: 100–240 VAC, 50/60 Hz, 42 VA (Safety Call II – Type CF with Functional Earth Conductor)
- Battery Life: At least 400 charge cycles; typically provides 10 hours of operation at 5.0 mL/hr using a 60 mL syringe
-
Operating Conditions:
- Temperature: 5° to 40° C (40° to 104° F)
- Relative Humidity: 20% to 95% (non-condensing)
- Ambient Pressure: 70 kPa to 106 kPa (10.2 to 15.4 psia)
-
Storage & Transportation Conditions:
- Temperature: –20° to 50° C (–4° to 122° F)
- Relative Humidity: 20% to 95% (non-condensing)
- Ambient Pressure: 70 kPa to 106 kPa (10.2 to 15.4 psia)
Reviews
There are no reviews yet.