Midmark Ritter M11 UltraClave Automatic Sterilizer
The Midmark Ritter M11 UltraClave is a robust, self-contained steam sterilizer tailored for high-volume sterilization needs in medical and dental offices, hospitals, clinics, nursing homes, and laboratories. This advanced sterilizer is designed to optimize sterilization processes efficiently, with a set of features aimed at reducing the time and effort needed for the sterilization of heat and moisture-stable reusable items.
Key Points
-
Intuitive Operation:
Features an intuitive display with simple prompts, making it easy to select from four pre-programmed sterilization cycles: Unwrapped, Pouched, Packs, and Handpieces. -
Safety and Efficiency:
Includes a Dry Cycle Door Stop that prevents the door from fully opening if slight residual pressure is still present in the chamber, enhancing safety and efficiency. -
Temperature and Pressure Monitoring:
Equipped with sensors to monitor chamber temperature and a pressure relief valve to manage steam pressure, ensuring safe and effective sterilization. -
Large Capacity:
Offers a spacious sterilization chamber with a diameter of 11 inches and a depth of 18 inches, accommodating large trays and a variety of instruments.
Specifications
-
Dimensions:
- Height (With Printer): 17.9 in. (45.2 cm)
- Width: 17.8 in. (45.2 cm)
- Depth (With Plug): 23.8 in. (60.5 cm)
- Counter Area Required: 17.8 x 21 in (45.2 x 53.3 cm)
-
Chamber Size:
- Diameter: 11.0 in (22.9 cm)
- Depth: 18 in (46 cm)
- Large Tray: 9 in x 15 in x 1 1/8 in (22.9 cm x 38 cm x 2.9 cm)
- Small Tray: 6 5/8 in x 15 in x 1 1/8 in (14.3 cm x 38 cm x 2.9 cm)
-
Weight:
- Empty: 99 lbs (44.9 kg)
- Full: 111 lbs (50.4 kg)
-
Water Reservoir Capacity:
- Approximately 1.4 Gallons (5.3 liters) to the full mark
- Usable Volume: 1 Gallon (3.8 liters)
-
Electrical Requirements:
- 115 VAC, 50/60 Hz, 15 Amps, Single Phase with a 1425 watt tubular immersion heater
- 230 VAC, 50/60 Hz, 10 Amps, Single Phase with a 1500 watt tubular immersion heater
-
Chamber Pressure:
- At 270 °F (132 °C): 27.1 psi (182 kPa)
- Safety Valve Setting: 40 psi (275.8 kPa)
-
Heat Emission:
- 5000 BTU/Hr during operation
-
Fuses:
- 115 VAC models: F1: 0.250 amp, 250 V, Slo-Blo; F2: 15 amp, 250 V, Fast-Acting
- 230 VAC models: F1: 0.125 amp, 250 V, Slo-Blo; F2: 8 amp, 250 V, Fast-Acting
-
Safety Standards:
- UL Listed to U.S. & Canadian Standards
- UL 61010-1, 2nd Edition; IEC 61010-2-040, 1st Edition
- CAN/CSA C22.2 No. 61010-1, 2nd Edition; CSA C22.2 No. 61010-2-040-07, Part 2-040, 1st Edition
- Meets ASME Boiler and Pressure Vessel Code requirements
Reviews
There are no reviews yet.