Steris Amsco Century V120 Steam Sterilizer
The Steris Amsco Century V120 steam sterilizer is a high-performance, mid-sized sterilization system designed to meet the needs of healthcare facilities. Available in both gravity and pre-vacuum configurations, this sterilizer provides flexibility in operation while maintaining high standards of sterilization efficiency. With a 20″ x 20″ x 38″ chamber, a color touchscreen display, and a foot pedal-activated vertical sliding door, the Century V120 offers ease of use, reliability, and superior sterilization performance.
Key Points
-
Flexible Configuration Options:
- Available in single or double door models, with options for gravity or pre-vacuum cycles, and can be mounted either open or recessed.
-
User-Friendly Controls:
- Equipped with a color touchscreen display and an integrated ink-on-paper printer for easy cycle tracking and documentation.
-
Automated Operation:
- Foot pedal-activated vertical sliding door allows for hands-free loading and unloading.
-
Water-Saving Design:
- Features electronic water-saving control with a resistance temperature detector condenser, optimizing water usage in the condensation process.
-
Advanced Drain System:
- Built-in steam purge assists in air removal and preheating, while the drain system prevents contaminants from entering the sterilizer and water supply.
Specifications
-
Overall Dimensions (Single Door, Enclosed):
- Height: 74.5 inches (189.2 cm)
- Width: 30 inches (76.2 cm)
- Depth: 45.125 inches (114.6 cm)
-
Chamber Size:
- Height: 20 inches (50.8 cm)
- Width: 20 inches (50.8 cm)
- Depth: 38 inches (96.5 cm)
-
Weight:
- Steam Model:
- 1,231 lbs (558 kg) – Single Door
- 1,606 lbs (728 kg) – Double Door
- Electric Model:
- 1,371 lbs (622 kg) – Single Door
- 1,726 lbs (782 kg) – Double Door
- Steam Model:
Sterilization Cycles
-
Flash Cycle (3-minute exposure)
- Sterilization Temp: 270°F (132°C)
- Sterilization Time: 3 minutes
- Dry Time: 1 minute
-
Flash Cycle (10-minute exposure)
- Sterilization Temp: 270°F (132°C)
- Sterilization Time: 10 minutes
- Dry Time: 1 minute
-
Express Cycle (Pre-Vac Only)
- Sterilization Temp: 270°F (132°C)
- Sterilization Time: 4 minutes
- Dry Time: 3 minutes
-
270°F Prevac Cycle (Pre-Vac Only)
- Sterilization Temp: 270°F (132°C)
- Sterilization Time: 4 minutes
- Dry Time: 20 minutes
-
275°F Prevac Cycle (Pre-Vac Only)
- Sterilization Temp: 275°F (135°C)
- Sterilization Time: 3 minutes
- Dry Time: 16 minutes
-
270°F Gravity Cycle
- Sterilization Temp: 270°F (132°C)
- Sterilization Time: 30 minutes
- Dry Time: 15 minutes
-
250°F Gravity Cycle (Fabric Packs)
- Sterilization Temp: 250°F (121°C)
- Sterilization Time: 15 minutes
- Dry Time: 30 minutes
Utility Requirements
Using House Steam:
- Steam: 1/2″ NPT, 50-80 psig (3.5-5.6 bar)
- Drain: 1-1/2″ ODT drain terminal
- Electrical – Controls: 120 Volt, 50/60 Hz, 1-phase, 2.0 Amps
- Sterilizer Feed Water: 1″ NPT, 30-50 psig (2.1-3.5 bar)
Using Steam Generator:
- Electrical – Controls: 120 Volt, 50/60 Hz, 1-phase, 9.5 Amps
- Electrical – Generator:
- 208V, 50/60 Hz, 3-phase, 83.2 Amps
- 240V, 50/60 Hz, 3-phase, 72.2 Amps
- 480V, 50/60 Hz, 3-phase, 37 Amps
- Sterilizer Feed Water: 1.0″ NPT, 30-50 psig (2.1-3.5 bar)
- Steam Generator Feed Water: 1/2″ NPT, 20-50 psig (1.4-3.5 bar)
Recommended Feed Water Quality
-
House Steam Requirements:
- Temperature: 40°-60°F (4°-16°C), max 70°F (21°C)
- Hardness: 50-120 mg/L (max 171 mg/L)
- Total Dissolved Solids: 100-200 mg/L (max 500 mg/L)
- pH: 6.8-7.5 (max 6.5-8.5)
- Total Silica: 0.1-1.0 mg/L (max 2.5 mg/L)
-
Steam Generator Requirements:
- Temperature: as supplied, max 140°F (60°C)
- Hardness: 0-17 mg/L (max 130 mg/L)
- Total Dissolved Solids: 50-150 mg/L (max 250 mg/L)
- pH: 6.8-7.5 (max 6.5-8.5)
- Total Silica: 0.1-1.0 mg/L (max 2.5 mg/L)
- Resistivity: 2000-6000 ohms/cm (max 26000)